Electrical conductivity (EC), total dissolved solids (TDS), and salinity are three closely related — yet distinct — water quality parameters. All three describe the ionic content of a water sample, but they are measured and reported differently. Confusing the three can lead to incorrect dosing in water treatment, failed product specifications, or misleading environmental reports. This guide explains what each parameter means, how they relate to one another, and how to choose the right meter and measurement approach for your industry.

1. Definitions

Electrical Conductivity (EC)

EC measures a water sample's ability to conduct an electrical current, expressed in µS/cm (microsiemens per centimeter) or mS/cm. It is a direct, real-time measurement — a conductivity meter applies a voltage between two electrodes and measures the resulting current. EC is temperature-dependent (typically referenced to 25 °C).

Total Dissolved Solids (TDS)

TDS represents the total concentration of dissolved inorganic and organic substances in water, expressed in mg/L (ppm). The definitive method is gravimetric (evaporate a known volume at 180 °C and weigh the residue), but in practice TDS is most often estimated from EC using a conversion factor.

Salinity

Salinity measures the concentration of dissolved salts — primarily NaCl — expressed in ppt (parts per thousand, ‰), PSU (practical salinity units), or g/L. In oceanography, salinity is derived from conductivity using the PSS-78 (Practical Salinity Scale 1978) algorithm. In industrial contexts, simpler EC-to-salinity conversion factors are used.

2. How They Relate

All three parameters originate from the same physical property — dissolved ions in solution:

  • EC → TDS: TDS (mg/L) ≈ EC (µS/cm) × conversion factor. The factor ranges from 0.5 to 0.7 depending on the dominant ions. A value of 0.5 is common for natural freshwater; 0.65–0.7 for mixed industrial water.
  • EC → Salinity: For NaCl-dominant solutions, salinity (ppt) ≈ EC (mS/cm) × 0.5 (rough estimate). Precise oceanographic salinity uses the PSS-78 polynomial.
  • TDS ≈ Salinity × 1,000 only when all dissolved solids are salts (not always the case in industrial wastewater).

Important: The EC-to-TDS factor is not universal. Using the wrong factor introduces systematic error. When precise TDS is needed, gravimetric analysis is the reference method.

3. Industry-Specific Guideline Values

ApplicationParameterTypical Limit
Drinking water (WHO)TDS< 600 mg/L (palatable); < 1,000 mg/L (acceptable)
Ultra-pure water (semiconductor)EC< 0.055 µS/cm (18.2 MΩ·cm resistivity)
Boiler feed waterEC / TDS< 1–10 µS/cm depending on pressure class
Irrigation waterEC< 700 µS/cm (sensitive crops); < 3,000 µS/cm (tolerant crops)
Aquaculture (freshwater)Salinity0–0.5 ppt
Aquaculture (marine)Salinity30–35 ppt
Food & beverage rinse waterEC< 5 µS/cm (final rinse verification)

4. Choosing the Right Meter

A modern conductivity meter can display all three parameters — EC, TDS, and salinity — from a single measurement, applying user-selectable conversion factors. Key selection criteria:

  • Cell constant: K = 0.1 for ultra-pure water; K = 1.0 for general water; K = 10.0 for high-concentration brine or wastewater.
  • Temperature compensation: Linear (simple) or non-linear (for accurate readings across wide temperature ranges).
  • TDS factor: Adjustable from 0.4 to 1.0 to match your specific water chemistry.
  • Form factor: Portable for field/plant use; benchtop for lab QC; online for continuous process monitoring.
  • Multi-parameter: Some meters combine EC/TDS/salinity with pH, DO, and turbidity in a single probe — reducing equipment count for field work.

5. Common Mistakes to Avoid

  • Using a fixed TDS factor for all sample types: Different water chemistries require different factors. Verify with gravimetric TDS if reporting for compliance.
  • Ignoring temperature: EC changes ~2 % per °C. Always use automatic temperature compensation (ATC) or report the reference temperature.
  • Comparing NTU and salinity: Turbidity (NTU) measures suspended particles; salinity measures dissolved salts. They are independent parameters.
  • Wrong cell constant: A K = 1.0 cell in ultra-pure water reads near zero with poor resolution. Match the cell to the expected range.

Explore Conductivity Meters

Sechang provides conductivity, TDS, and salinity meters for laboratory, portable, and online applications. Our engineers can help you select the right cell constant, conversion factor, and integration approach for your specific water quality program.

8. Industry-Specific Guidance

8-1. Semiconductor / Electronics Manufacturing

Ultrapure water (UPW) for semiconductor fabs requires resistivity ≥ 18.0 MΩ·cm, equivalent to conductivity ≤ 0.056 µS/cm. At this level, TDS is immeasurably low. The correct parameter is resistivity or conductivity — not TDS. TDS sensors are not appropriate for UPW quality monitoring. Use dedicated resistivity/conductivity cells certified for high-purity water (ASTM D5391, SEMI F63).

8-2. Food and Beverage / Brewing

Brewing water chemistry is characterized by specific ion concentrations (calcium, magnesium, bicarbonate, sulfate, chloride) rather than bulk TDS. While TDS gives a rough mineral content indication, brewers use complete water analysis reports for recipe design. Conductivity is used as a quick in-process check for cleaning verification (high conductivity = CIP chemical present) and water softener breakthrough detection.

8-3. Aquaculture

In marine and brackish aquaculture, salinity (reported as PSU or g/kg) is the ecologically relevant parameter — not TDS or conductivity directly. Modern aquaculture meters calculate salinity from conductivity using the Practical Salinity Scale (PSS-78). Always enter temperature accurately because the conductivity-to-salinity conversion is temperature-sensitive at marine concentrations.

8-4. Reverse Osmosis and Membrane Systems

RO system operators use conductivity rejection percentage as a key performance indicator: Conductivity Rejection (%) = (1 − Permeate EC / Feed EC) × 100. A well-performing RO system typically achieves 95–99 % conductivity rejection. Declining rejection signals membrane fouling, scaling, or mechanical damage. TDS meters can substitute for conductivity in this application, but conductivity gives faster response for inline monitoring.

9. Measurement Accuracy Factors

Several factors affect the accuracy of conductivity, TDS, and salinity measurements:

  • Temperature: Conductivity increases approximately 2 % per °C. Always use automatic temperature compensation (ATC) referenced to 25 °C for comparable results.
  • Cell constant: Conductivity cells are characterized by a cell constant (K, in cm⁻¹). Low K values (0.01–0.1) suit low-conductivity UPW; high K values (1–10) suit high-conductivity industrial process water. Using the wrong cell constant causes systematic errors.
  • Electrode fouling: Mineral scale, oil films, and biological growth increase apparent resistance, causing artificially low conductivity readings. Clean electrodes with mild acid (for scale) or appropriate solvent (for organics).
  • TDS conversion factor: Different water types have different dominant ion species. The conversion factor from µS/cm to mg/L varies from about 0.5 (NaCl-dominated) to 0.9 (mixed ion) to 1.0 (varies). Confirm which factor your meter applies and whether it matches your sample type.

10. Choosing a Conductivity/TDS Meter: Key Specifications

SpecificationWhat to Look For
Measurement rangeCovers your application: 0–200 µS/cm (drinking water), 0–200 mS/cm (seawater/industrial)
Resolution0.001 µS/cm for UPW; 0.01 µS/cm for drinking water; 0.1–1 mS/cm for industrial
Accuracy±0.5–1 % of reading for quality instruments
Temperature compensationAutomatic (ATC) with configurable temperature coefficient (α)
Cell constant (K)Matches your conductivity range; interchangeable cells for multi-range use
Salinity calculationPSS-78 for marine; linear NaCl for general
TDS factorUser-adjustable 0.4–1.0 for different water types
Data loggingGLP record keeping with timestamp and operator ID for compliance

▶ Browse Conductivity Meters & Request a Quote

11. Summary: Matching Parameter to Purpose

The following decision tree summarizes when to choose each parameter:

  • Conductivity (EC): Use when you need a universal, temperature-correctable measure of dissolved ion load. Best for process control, RO rejection, water softener monitoring, and regulatory compliance monitoring.
  • TDS: Use when communicating with non-technical audiences who understand "dissolved solids" in drinking water context. Acceptable for irrigation and consumer water quality assessment. Avoid as a sole parameter in research or compliance applications.
  • Salinity: Use in marine, estuarine, or high-concentration brine applications where specific gravity or osmotic pressure is relevant to the process or organism being monitored.

Remember: conductivity is the fundamental measurement; TDS and salinity are both derived values calculated from conductivity. When in doubt, report conductivity — it is the unambiguous, standardized value that any downstream user can convert to their preferred derived parameter.

Sechang carries conductivity, TDS, and salinity meters for all application tiers — from low-cost pocket testers for field use to high-precision inline analyzers for semiconductor and pharmaceutical water systems. Our team can recommend the right cell constant and instrument platform for your specific conductivity range and accuracy requirements.

▶ Browse Conductivity / TDS Meters — Get a Quote

12. Verifying Conductivity Meter Performance in Your Application

Before finalizing a conductivity/TDS meter for a critical application, verify performance with your actual water sample rather than manufacturer-supplied reference solutions alone. Test calibration stability by measuring a known reference standard at the start and end of a full work shift — drift exceeding the specification indicates either a cell constant issue or an inadequate stabilization time. For online applications, perform a comparison between the online sensor and a benchtop reference meter on the same sample at least weekly for the first month of operation, until the sensor's long-term stability in your process environment is established. Document all comparison data as part of your instrument qualification record. For regulated applications, retain these records for the full retention period required by your quality management system.