When measuring noise, a sound-level meter applies a frequency weighting matched to the ear's response or to the measurement purpose. The two most widely used are A-weighting and C-weighting; using a weighting that does not match the purpose can make the reading itself non-compliant with the applicable regulation. This guide sets out the difference between the two and which to choose by measurement purpose and regulation.
1. What Frequency Weighting Is
The human ear does not hear all frequencies equally. A sound-level meter's frequency weighting applies a different correction at each frequency to reflect this hearing response or the measurement purpose. For the same sound, the measured decibel (dB) value changes depending on which weighting is applied.
2. A-Weighting vs C-Weighting
A-weighting (dBA) strongly attenuates low frequencies, to which the ear is relatively insensitive, approximating the loudness people actually perceive. Most environmental and workplace noise regulations are based on A-weighting. C-weighting (dBC) barely attenuates low frequencies, reflecting a wide frequency band relatively flat. It suits noise with strong low-frequency content and peak sound-pressure evaluation.
3. Choosing by Measurement Purpose
- Occupational / environmental noise-regulation assessment: mostly A-weighting (dBA) — human-impact basis
- Peak (impulse) noise evaluation: C-weighting peak (dBC peak) — managing instantaneous maximum sound pressure
- Low-frequency / machinery vibratory noise: use C-weighting to check the low-frequency contribution
- Difference of A and C (dBC−dBA): used as an indicator of how much low-frequency content is present
4. Weighting Required by Regulation
Hearing-conservation and workplace-noise criteria in industrial safety, and environmental-noise regulations, typically require the A-weighted equivalent level (LAeq). Impulse-noise or peak-management items, on the other hand, may require a C-weighted peak. Before measuring, first confirm the weighting and time weighting (Fast/Slow/Impulse) that the applicable regulation requires.
5. Selection Checklist
- Does the applicable regulation call for A-weighting or C-weighting?
- Is the target the equivalent level or the peak sound pressure?
- Is the noise dominated by low-frequency content?
- What time weighting is required (Fast/Slow/Impulse)?
- Does the instrument support the weighting, and is it calibrated?